In brief
Based on the essays of Ezhil Ganesan (MU Plovdiv), improved and checked by AceMed
- Monitoring means measuring, recording and interpreting vital functions to detect problems early; the monitor never replaces the anaesthetist's eyes, hands and judgement, and trends matter more than single values.
- Minimum standard for every anaesthetic, including sedation: continuous ECG, SpO₂ (pulse oximetry), end-tidal CO₂ (capnography) and non-invasive blood pressure every 3–5 min, plus inspired O₂, volatile agent concentration, airway pressure, neuromuscular monitoring (if relaxants) and temperature.
- The ICU adds invasive and advanced tools (arterial line, CVP, cardiac output devices, ICP, EEG, continuous core temperature) according to the patient's illness and the question being asked.
Haemodynamic monitoring
Haemodynamic monitoring measures pressure, flow and oxygen delivery in the cardiovascular system. Start with clinical signs (pulse, capillary refill, skin, urine output, mentation) and add devices as the risk increases.
Electrocardiography (ECG)
- Indicated for every patient under anaesthesia (no contraindication); detects rate, rhythm, conduction and, with a 5-lead system (leads II and V₅), ischaemia.
- Waves: P atrial depolarisation; QRS ventricular depolarisation; ST segment the plateau between depolarisation and repolarisation (elevation or depression suggests ischaemia); T ventricular repolarisation; PR interval atrial to ventricular conduction time; QT interval total ventricular depolarisation and repolarisation.
Arterial blood pressure
- Pressure is pulsatile: systolic (SBP), diastolic (DBP), pulse pressure = SBP − DBP, and mean arterial pressure MAP ≈ (SBP + 2 × DBP) / 3. MAP is the pressure that drives organ perfusion; a usual aim is ≥ 65 mmHg. Peripheral (radial) systolic pressure is higher than aortic because of wave reflection.
Non-invasive blood pressure (NIBP)
- Indication: any anaesthetic; measure every 3–5 min (more often in unstable patients). Contraindication: do not place the cuff on a limb with a dialysis fistula, vascular disease, IV line or lymphoedema/mastectomy side.
- Palpation: inflate until the pulse vanishes, release slowly; gives only systolic pressure.
- Auscultation: Korotkoff sounds; the first sound is systolic and disappearance of sound is diastolic.
- Oscillometry: the standard automated method; the device measures oscillation amplitude, derives mean pressure (maximal oscillation) and calculates SBP and DBP. It is unreliable in arrhythmia (atrial fibrillation), shivering and shock, and a wrongly sized cuff gives false values (too small reads high; too large reads low; the cuff width should be about 40% of arm circumference).
- Doppler: useful in obese, paediatric or shocked patients when other methods fail.
- Arterial tonometry and finger-cuff methods: beat-to-beat curves without puncture; sensitive to movement and need calibration.