Based on the essays of Ezhil Ganesan (MU Plovdiv), improved and checked by AceMed
- The anaesthesia workstation takes medical gases from the supply, lowers and controls their pressure and flow, adds a vapour, and delivers the fresh gas to a breathing system joined to the patient's airway, with a ventilator, monitors, alarms and waste-gas scavenging built around it.
- Its safety design exists to prevent delivery of a hypoxic mixture: oxygen failure alarm, nitrous oxide cut-off, an oxygen-to-nitrous-oxide ratio link and agent-specific vaporisers.
- Breathing systems are either Mapleson A–F (no absorber, fresh-gas-flow dependent) or the circle system (valves plus CO₂ absorber, economical at low flow); the Bain (D) and Jackson–Rees (F) are the ones to know.
Anaesthesia workstation – basic components
A workstation is the anaesthesia machine together with its ventilator, breathing system, monitoring, alarms and scavenging, built as one integrated unit.
Route of the gases: supply, regulators, flowmeters, vaporiser, common gas outlet, breathing system, patient
Gas supply
- Pipeline: oxygen, medical air and nitrous oxide arrive from the hospital supply at about 4 bar (≈400 kPa) through non-interchangeable, colour-coded connectors.
- Cylinders are the back-up. Each yoke uses the pin-index system, so a cylinder fits only its own gas. An oxygen cylinder holds compressed gas (about 137 bar when full in a standard E-size cylinder), so its gauge pressure shows the contents; a nitrous oxide cylinder holds a liquid, so the pressure stays near 44–50 bar (depending on temperature and country) until it is almost empty, and the weight shows the contents.
- Regulators reduce cylinder pressure to a few bar, a safe working level for the flowmeters.
- Oxygen keeps the patient oxygenated and drives the pneumatic alarm; medical air dilutes oxygen and avoids nitrous oxide; nitrous oxide is a weak anaesthetic and analgesic carrier gas, now used less.
Gas mixing and flow control
- Flowmeters (variable-orifice, tapered tubes with a bobbin or electronic displays) set the flow of each gas in L/min; the mixed gas is the fresh gas flow (FGF).
- Oxygen safeguards: a low-pressure alarm sounds if oxygen pressure falls; a fail-safe valve cuts off nitrous oxide (and other gases) when oxygen fails; a linked hypoxic guard keeps oxygen at least 25% of the oxygen/nitrous oxide mixture (maximum nitrous oxide : oxygen ratio about 3 : 1). Oxygen is always the last gas to enter the common manifold, so a leak upstream cannot dilute it.