Introduction
Besides the scalpel and the rotary bur, oral surgery uses three energy-based ways to cut and coagulate: high-frequency current, ultrasonic (piezoelectric) vibration and laser light. Each works on a different physical principle, so each fits a different job: soft tissue with haemostasis, selective bone cutting, or work targeted by wavelength.
High-frequency electrosurgery: principle and equipment
Electrosurgery uses radiofrequency alternating current, far above 100 kHz (dental units work in the low-megahertz range). At this frequency nerves and muscles are not stimulated (no faradic effect). Heat forms inside the tissue from its electrical resistance; the electrode itself stays cold.
- Monopolar: an active electrode plus a dispersive (passive) return plate or antenna.
- Bipolar: current flows only between the two tips of a forceps. It is mainly used for haemostasis and gives less current spread.
Waveforms in classic dental teaching:
- fully rectified, filtered (continuous): pure cut with little lateral heat;
- fully rectified: cut plus some coagulation;
- partially rectified: coagulation;
- spark-gap (highly damped): fulguration.
The techniques are electrosection (electrotomy), electrocoagulation, fulguration and electrodesiccation. Electrodes are chosen by job: fine wire or needle for incision, loop for planing and excision, ball for coagulation. The term radiosurgery is used for high-frequency (MHz) units that cut with minimal lateral heat.
Working rules:
- keep the field moist but not flooded, and use high-volume suction to clear the smoke plume;
- use light, quick, pressure-free strokes and let the tissue cool for a few seconds between passes over the same spot;
- never touch bone, periosteum, metal restorations or implants;
- no flammable agents (alcohol, oxygen-enriched air).